GasAlertMax
User Manual
To change the factory-set alarm setpoints, refer to the
Calibration and Setting Alarm Setpoints section.
Note
You can disable an alarm by setting the alarm
setpoint to 0 (display: - -).
Stopping a Gas Alarm
The low, high, and STEL alarm stops when the ambient gas
level returns to the acceptable range.
The detector computes the TWA value based on an 8-hour
workday. Only deactivating the detector can stop the TWA
alarm.
Acknowledge Latched Alarm
If the latched alarm option is activated, the audible and visual
alarms continue to sound and flash until the low, high, or
STEL alarm condition is acknowledged. Press C to
deactivate the audible and visual alarms when the current
ambient gas reading falls below the low alarm level. The
alarms cannot be deactivated if an alarm condition is still
present. A TWA alarm condition will not reset.
Note
A STEL alarm may take up to 15 minutes of clean
air readings before the latched alarm can be
deactivated.
Sensor Alarm
The detector tests for a missing or defective sensor during
the activation self-test. Refer to the
Troubleshooting section.
Pump Alarms
The internal pump draws air over the sensors. If the pump
stops working or the pump filter becomes clogged, the
detector activates the pump alarm.
Pump Alarm
Deactivate the detector immediately
and clear the pump inlet, hose, and
filter.
Blocked Pump Alarm
If the pump alarm continues and the
LCD advises the inlet is blocked, see
the
Pump Calibration section.